Chapter 2 Installation Of The Device; Environmental Conditions; Requirements - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Installation of the device
Table of Contents

"Environmental Conditions"

"Room ventilation"
"Space requirements"
"Transport"
"Stacking"
"Retrofitting/Modifications"
Environmental Conditions
The device must only be operated at locations that meet the particular ambient conditions listed
below:

Requirements:

• Draft-free and dry location.
• The minimal distance to adjacent surfaces must be observed on all sides, see
on
page
• The operating room must be equipped with appropriate room ventilation.
• Solid, level, fire-proof surface.
• Vibration-proof substructure (floor stand, lab table) capable of bearing the weight of the device
and of accessories (particularly if several devices are stacked).
• The electrical system of the device has been designed for an operating height of up to 2000 m
above sea level.
• To ensure a constant incubation temperature of 37 °C, the ambient temperature must be within a
range of +18 °C to +34 °C.
• Relative humidity up to 80% (max.).
• No direct exposure to sunlight.
• Devices that produce excessive heat must not be placed near the location of S
S
TERI-CYCLE
Thermo Scientific
on
on
page 2-2
on
page 2-2
on
page 2-3
on
page 2-3
on
2-2.
i250
page 2-1
page 2-6
2
"Space requirements"
i160 /
TERI-CYCLE
i160 / i250
S
TERI-CYCLE
2-1