Overview of event entries, Part II:
Code
0x52
0x53
0x60
0x61
0x62
0x63
0x70
0x71
0x72
0x90
0x91
0xe0
0xff
Examples of data logger codes
An entry in the data logger is 16 bytes large and has the following structure:
1st byte: indicates the event (e.g. door open 1x31, measured value entry 0x01)
2nd byte: day of entry
3rd byte: month
4th byte: year
5th byte: hours
6th byte: minutes
7th and 8th byte: device status
9th to 16th byte: various data on the event
Thermo Scientific
Event
auto-start completed with error
auto-start stopped manually
Start steri-run
steri-run successfully completed
steri-run completed with error
steri-run stopped manually
Gas mon. cylinder A empty
Gas mon. cylinder B empty
Gas mon. changeover
Start low humidity
Stop low humidity
Delete data logger
Last entry of data logger
12
Annex: Data communication
Examples of data logger codes
Special information (Byte 8-15)
Status / error registration for
temperature, CO
, O
and rH
2
2
Status / error registration for
temperature, CO
, O
and rH
2
2
Status / error registration for
temperature, CO
, O
and rH
2
2
Current actual values for
temperature, CO
, O
and rH
2
2
Status / error registration for
temperature, CO
, O
and rH
2
2
Status / error registration for
temperature, CO
, O
and rH
2
2
Gas monitoring status (2 bytes),
4 bytes empty
Gas monitoring status (2 bytes),
4 bytes empty
Gas monitoring status (2 bytes),
4 bytes empty
Current actual values for
temperature, CO
, O
and rH
2
2
Current actual values for
temperature, CO
, O
and rH
2
2
Current actual values for
temperature, CO
, O
and rH
2
2
No information, not even on date,
time, and status
i160 / i250
S
TERI-CYCLE
12-15