Wipe/spray disinfection

The manual wipe/spray disinfection is carried out in three stages:
• Predisinfection
• Cleaning,
• Final disinfection
CAUTION
• Alcoholic disinfectants!
Disinfectants having an alcohol content of more than 10% may form, in combination with air,
easily combustible and explosive gas mixtures.
When using such disinfectants, avoid open flames or exposure to excessive heat during the
entire disinfection process!
• Use such disinfectants only in adequately ventilated rooms.
• After the disinfectant has been allowed to react, wipe the cleaned device components
thoroughly dry.
• Observe safety regulations to avoid fire and explosion hazard caused by alcohol-containing
disinfectants.
CAUTION
Chloride-containing disinfectants!
Chloride-containing disinfectants can corrode stainless steel.
• Use only disinfectants that do not affect stainless steel!
WARNING Electric shock!
Contact with current-carrying components may cause a lethal electric shock.
Prior to cleaning and disinfection work, disconnect the device from the power supply!
• Turn the device off using the power switch.
• Unplug the power connector and protect it against accidental reconnection.
• Make sure the device is de-energized.
Thermo Scientific
8
Cleaning and disinfection
Wipe/spray disinfection
i160 / i250
S
TERI-CYCLE
8-3