Retrofitting/Modifications - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

2
Installation of the device

Retrofitting/Modifications

The devices are secured to one another by their own weight.
.
CAUTION Transporting stacked devices!
Stacking elements are no connection elements. Stacked devices provided with a mobile base rack
must only be moved inside rooms on even floors without any slopes.
Note Fastening on mobile racks:
If the devices are installed onto mobile racks, make sure that the rollers are secured with a locking
brake during the operation of the incubators and that the rollers are oriented toward the front for
increased stability.
Condensate formation during the operation of stacked devices:
As a rule when operating stacked devices of type S
be placed between the two devices for thermal separation.
If stacked devices are operated at an ambient temperature of more than 28 °C, an overtemperature
alarm is triggered on the upper device while the steri-run decontamination routine is run for the
lower device. Condensate formation may then occur on the upper device.
Retrofitting/Modifications
The following components can be retrofitted to the standard version:
S
TERI-CYCLE
• threefold gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• infrared (IR) CO
• 4-20 mA data interface.
• gas cylinder monitor
S
TERI-CYCLE
• threefold gas-tight screen (the gas-tight screen is installed in place of the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• split inserts,
• infrared (IR) CO
• 4-20 mA data interface.
• gas cylinder monitor
Note Modifications:
Retrofittings and modifications must only be performed by the Technical Service of Thermo
Electron LED GmbH.
i160 / i250
S
2-6
TERI-CYCLE
i160
sensor (replaces the standard TCD sensor),
2
i250
sensor (replaces the standard TCD sensor),
2
i160 and i250, an adapter plate should
TERI-CYCLE
Thermo Scientific