Thermo Scientific STERI-CYCLE i160 User Manual page 89

Co2 incubator
Table of Contents

Advertisement

Notes on using the auto-start routine:
To ensure that the specified accuracy of the CO
always be started using the auto-start routine if:
• a difference of more than 1 °C is entered upon setting the temperature set value,
• the Low Humidity function is enabled/disabled,
• the device is restarted after an extended interruption of operation.
The auto-start routine should be run at least every three months on the occasion of cleaning and
maintenance works.
Duration of the auto-start routine:
Running the auto-start-routine usually takes 5 to 7 hours. At low room temperatures and when the
device is cold, it may take up to 10 hours until the auto-start routine has been completed. If the glass
door is opened or if the power supply of the device is interrupted while the auto-start routine is
running, the routine is interrupted and rerun after the glass door has been closed or the power supply
has been reestablished.
Start conditions for the auto-start routine:
Prior to running the auto-start routine, set the CO
and make sure that the atmosphere in the work space only consists of ambient air. The water reservoir
in the workspace must be filled sufficiently with water.
Inhibition of the auto-start routine:
The auto-start routine will not start if any of the following error conditions exists.
Temperature control loop:
• Sensor breakage,
• Actual value high, exceeding set value (excessive deviation),
• Actual value low, falling short of set value (excessive deviation),
• Actual value not plausible
• calibration values too high or too low,
• No communication with sensor.
Gas supply CO
• No communication with sensor.
In this case, the auto-start key is dimmed and its function is not available.
Thermo Scientific
control loop:
2
measuring system is maintained, the device should
2
and O
set values to the desired values
2
2
S
TERI-CYCLE
6
Handling and control
auto-start function
i160 / i250
6-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Steri-cycle i250

Table of Contents