Figure 9-1. Preparing Temperature Calibration

Comparison measurement procedure:
1. Turn on the device using the power switch.
2. Set the temperature set value and allow the device to heat up. This may take several hours.
3. Place the measuring instrument (2) onto the center area of the workspace.
Alternatively, a temperature sensor may be positioned in this location. Route the connecting cable
either through the measurement opening in the glass door (3) or through the access port (1) in the
rear panel of the device.
4. Close the doors.
5. Wait until the temperature value displayed on the measuring instrument has stabilized.

6. Temperature calibration procedure

Temperature Calibration Procedure
Measurement example:
• Temperature set value: 37 °C
Reference temperature: 36.4 °C
1. Press the Temperature Display key.
• The Temperature menu
Thermo Scientific
(Figure
9-2) is displayed.
9
Maintenance
Temperature Calibration Procedure
i160 / i250
S
TERI-CYCLE
9-3