Factory presettings of the iCan™ Touchscreen controls
Upon delivery of the device, the following set values have been preset:
• Temperature: 37 °C
• CO
• O
-content (optional): 21,0 %
2
Note CO
Since the CO
monitoring system are disabled at a set value of 0%.
Since the O
system are disabled at a set value of 21%.

Heat-up phase of the control loop sensors

When the device has been switched on, the control loop sensors pass through a heat-up phase of
varying duration during the start process:
1. Temperature control loop: 10 s
2. CO
3. O
2

Figure 6-6. Heat-up phase display

The start process is indicated by an audible signal. During the heat-up phase, the displays show dots
(...) instead of values:
• Temperature display,
• CO
• O
2
Thermo Scientific
-content: 5,0 %
2
-/O
control:
2
2
concentration of the air is nearly 0%, the CO
2
concentration of the air is 21%, the O
2
sensor heat-up period (TCD and IR): 5 min
2
control loop: 5 min
display and
2
display field
Factory presettings of the iCan™ Touchscreen controls
control and the control loop error
2
control and the control loop error monitoring
2
S
TERI-CYCLE
6
Handling and control
i160 / i250