Cleaning and disinfection
Table of Contents

"Cleaning,"

"Decontamination procedures"
"Preparing for disinfection or steri-run"
"Wipe/spray disinfection"
"Steri-run decontamination routine"
Cleaning,
CAUTION Incompatible cleaners!
Some device components are made of plastic. Solvents can dissolve plastics. Strong acids or caustic
solutions can cause to become brittle of the plastic. For cleaning plastic components and surfaces,
do not use solvents that contain hydrocarbons, solvents with an alcohol content of more than 10%
or strong acids or caustic solutions!
Moisture-sensitive components!
Do not spray cleaning agent onto the touchscreen and the control box at the rear of the device.
When wiping the device clean, always make sure that moisture does not enter into these
components.
Cleaning exterior surfaces:
1. Thoroughly remove dirt residues and deposits using a solution of tepid water and dishwashing
agent.
2. Wipe the surfaces clean using a clean cloth and clear water.
3. Then, wipe the surfaces dry using a clean cloth.
Cleaning display:
CAUTION Moisture-sensitive display!
Do not spray or wipe the display with cleaner.
• Clean display using a dry cloth of 100% micro fiber!