• The selection dialog shown in
3. Select the menu command Gas tight screen.
• The input dialog shown in

Figure 6-39. Gas tight screen setting

1. To toggle between two options:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the Enter key.
• The system returns to the Options menu.
Activating/de-activating the water level sensor
For incubation operation with ambient humidity or if the auto-start routine is to be run dry (without
water), the water level sensor can be switched off. This prevents alarm messages for the water level
sensor by the device-integral monitoring system from being issued:
WARNING When the water level sensor is de-activated, the steri-run routine can be started
despite of the presence of a water fill in the water reservoir. This is improper usage and may destroy
the fan motor.
Activating/de-activating the water level sensor
1. Press the Menu key.
2. Select the menu command Options.
• The selection dialog shown in
3. Select the menu command Water level sensor.
• The input dialog shown in
Thermo Scientific
Figure 6-35
is displayed.
Figure 6-39
is displayed.
Figure 6-35
is displayed.
Figure 6-40
is displayed.
6
Handling and control
User configuration
i160 / i250
S
TERI-CYCLE
6-35