Alarm Contact; Power Supply Connection - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

3
Description of the device
Supply interface
I
= 4m A,
0
I
= 20m A, MB = measurement range
max
The meaning of the output signals is explained in the table below:
Current
4 mA
20 mA
2 mA
0 mA
The four 4-20 mA output signals are broken out onto a detachable, 8-pin terminal block (3,5 mm /
0.14 inch). The mating connector is delivered as a standard accessory. The pin-out can be taken from
the following table.
Pin No.

Alarm contact:

The device can be connected to an on-site, external alarm system (such as a private branch telephone
exchange, a facility monitoring system, visual or audible alarm indicators.
For this purpose, a potential-free alarm contact is preinstalled in the device. This contact is broken out
to the control box on the backside of the unit
Note Alarm contact:
The alarm contact responds for all errors reported by the control loops (see chapter Error
Messages).

Power supply connection:

Mains power is connected to the unit using a cable with an IEC connector plugged into the IEC socket
in the control box
accessible at all times.
i160 / i250
S
3-16
TERI-CYCLE
MV = measured value (°C, % or rH%)
The measured value is smaller than or equal to the lower boundary of the
measurement range.
The measured value is greater than or equal to the upper boundary of the
measurement range.
The unit is fully functional, but there is no valid measured value available (for
example, the optional sensor is not installed, the unit is still in the warm-up
phase, or similar.).
Error
Channel Description
1
Channel 1: Temperature
2
Channel 1 Ground
3
Channel 2: CO2
4
Channel 2 Ground
5
Channel 3: O2
6
Channel 3 Ground
7
Channel 4: Not used
8
Channel 4 Ground
(7/Figure
3-9). The mains socket must be easily identifiable by the user and freely
Significance
(6/Figure
3-9).