Gas Connection; Label; Usb Interface; 4-20 Ma Interface - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Gas connection:

The gas supply lines between the device and the gas supply system are connected using the supplied
connecting hoses. CO
3/Figure 3-9
All process gases must be supplied to the device at a fixed pressure that has been preset within a range of
0.8 to 1.0 bar (11.6 to 14.5 psi) and must remain unchanged.
Before the gases are fed into the workspace, they flow through a gas inlet filter with a separation rate of
99.998 % for a particle size of 0.3 µm (HEPA-filter quality).
The illustration shows the optional combined gas connection without the optional gas monitoring
system.

Label:

The label
notes about the electrical fusing of the device.

USB interface:

The USB interface
1.1 / USB 2.0 / USB 3.0 full speed compatible - allows the fast (and temporary) acquisition of the
major operating parameters (temperature, CO

4-20 mA interface:

The optional 4-20 mA interface
temperature, CO
output current of 4-20 mA each. This is accomplished by a microprocessor supported by a 4-channel
D/A converter. The D/A converter provides a resolution of 16 bit per channel, equivalent to 65536
incremental values.
The interface port accepts various customer-supplied external measurement devices with 4-20 mA
inputs - for example, the Thermo Scientific Smart-Vue Wireless Monitoring System (not supplied).
The table below lists the output signals:
Channel
No.
1
2
3
4
Parameters marked with an asterisk "*" are only output if the respective option is installed.
The measured samples are governed by the following function:
MW
Thermo Scientific
and O
/N
2
2
2
as well as
3/Figure
3-10).
(4/Figure
3-9) contains information about gas supply, an alarm contact terminal legend, and
(5/Figure
3-9) allows for connecting the incubator to a PC. This connection - USB
(2/Figure
concentration, O
2
Output Parameter (actual
value)
Temperature
CO
concentration
2
O
- concentration*
2
Not used
I
I
max
0
------------------- -
=
+
I
I
×
0
MB
are supplied to the unit through dedicated connectors (2 and
-/O
-/N
concentration, error codes, etc.).
2
2
2
3-10) converts the digital actual values displayed for
concentration (optional) and relative humidity into a continuous
2
Resolution
0.08 MA/°C
0.8 mA/%
0.16 mA/%
3
Description of the device
Supply interface
Measuring
Output
range
signal
0...200 °C
4...20mA
0...20% CO
4...20mA
2
0...100% O
4...20mA
2
i160 / i250
S
TERI-CYCLE
3-15