Troubleshooting - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

6
Handling and control
Measures after power outage

Troubleshooting

The error tables indicate source of error, cause of error, and possible corrections.
For any communication with the Technical Service of Thermo Fisher Scientific, please have the device
data ready.
Control
Failure message
loop
System
Device door open too
long
Error: Display
Error: EEPROM main
board
Error: data logger
Error: steri-run
Power down during
steri-run
Error: auto-start
Error: ADC
Error: Fan
IR sensor changed
i160 / i250
S
6-48
TERI-CYCLE
Cause
Device door has been
open for longer than 10
minutes
Display not
communicating with
*1)
the main board
EEPROM on main
board is defective
Error while writing to
the memory of the data
logger. Incubator is still
operative.
Error in Steri-Run
routine
Power failure during
Steri-Run routine
Error in auto-start
routine
Measurement of
reference resistance out
of tolerance
Actual value of the fan
is out of tolerance.
New serial number
detected
Alarm
Remedy
relay
Close device door
X
Reset unit. If the
X
error occurs
repeatedly, call
service.
Reset unit. If the error
X
occurs repeatedly, call
service.
Reset unit. If the error
occurs repeatedly, call
service.
Reset unit. If the error
X
occurs repeatedly, call
service.
Restart unit and
X
re-launch Steri-Run.
Re-launch auto-start.
X
If the error occurs
repeatedly, call ser-
vice.
Reset unit. If the error
x
occurs repeatedly, call
service.
Reset unit. If the error
X
occurs repeatedly, call
service.
Acknowledge the
X
alarm
Audible
Log
alarm
X
X
X
X
X
X
X
X
X
X
X
X
x
x
X
X
X
X
Thermo Scientific