Figure 6-37. Alarm Relay Setting - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Figure 6-37. Alarm relay setting

1. To toggle between two states:
• Press the + key.
or
• Press the - key.
2. To accept and save the change:
• Press the Enter key.
• The system returns to the Options menu.
Low humidity setting
If condensation occurs on the culture containers due to high relative humidity, the humidity in the
workspace can be set to a lower level. At the factory, the device control is preset to high humidity
(approx. 93% relative humidity.
The relative humidity in the workspace is lowered from approx. 93% to approx. 90%.
The modification requires an extended adaption phase. To ensure that it effectively prevents dew
formation on culture containers, it must be used as a permanent setting.
Reducing humidity in the workspace
1. Press the Menu key.
2. Select the menu command Options.
• The selection dialog shown in
3. Select the menu command Low humidity.
• The input dialog shown in
Thermo Scientific
Figure 6-35
is displayed.
Figure 6-38
is displayed.
6
Handling and control
User configuration