Chapter 5 Operation; Preparing The Device - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

Operation
Table of Contents

"Preparing the device"

"Starting operation"
Preparing the device
The device must not be released for operation before all major start-up activities (see
page
4-1) have been completed.
Device check:
Prior to starting operation, the following device components must be checked for their correct function
• Gas hoses must be seated tight on the connecting filter and must be secured using a hose clamp.
• The access port must be capped.
• The pressure compensation opening must be permeable, its insert must be installed in the
workspace.
• The glass door seal must not be damaged.
• The glass door/gas-tight screen measurement opening must be capped.
• The shelf system components must be installed safely.
• The airbox with the HEPA-filter and the air duct must be installed according to the instructions.
Decontaminating the device workspace:
• Run the steri-run decontamination routine (see
or decontaminate the workspace according to the hygiene regulations set forth by the operator.