Shelf System; Figure 3-16. Shelf System Components - Thermo Scientific STERI-CYCLE i160 User Manual

Co2 incubator
Table of Contents

Advertisement

3
Description of the device
Workspace Components
Note Operating conditions:
When accessories are to be operated in the workspace of the CO
requirements must be observed (see table). The energy introduced into the workspace affects the
beginning of the temperature control range. When additional heating sources are introduced into
the workspace, condensation (e.g. at the glass door) may occur.
Introduced energy
0 W
5 W
10 W
15 W
20 W
*RT = Room temperature

Shelf system

1
2
3

Figure 3-16. Shelf system components

The support rails
the support hooks
shelves
(1/Figure
come equipped with a single-piece bottom insert, as shown on the left side in
i160 with a cutout for refilling
page
4-1.
i160 / i250
S
3-22
TERI-CYCLE
Beginning of the temperature control range
General
RT + 3 °C
RT + 6.5 °C
RT + 9.5 °C
RT + 13 °C
RT + 16 °C
(3/Figure
3-16) of the shelf system are perforated every 42 mm (1.66 inch) so that
(2/Figure
3-16) can be inserted variably for any culture container size required. The
3-16) have an integrated tilt protection and withdrawal stop. S
(4/Figure
3-16). For details on using the shelf system, see
incubator, the ambient condition
2
Example:
RT* = 21 °C
24 °C
27.5 °C
30.5 °C
34 °C
37 °C
4
i250 models
TERI-CYCLE
Figure
3-16, S
TERI-CYCLE
"Start-up"
on
Thermo Scientific