Power supply connection

Figure 4-14. Power supply connection

WARNING Electric shock!
Contact with current-carrying components may cause a lethal electric shock.
Before connecting the device to the power supply, check plug and power supply cable for damage.
Do not use damaged components when connecting the device to the power supply!
Connect the unit to a correctly installed and grounded power supply source fused with a B 16 circuit
breaker.
Connection to the power supply system:
1. Before connecting the device to the power supply, check to see if the voltage of the power outlet
corresponds with the specifications on the nameplate to the left of the power switch. If the ratings
given for voltage (V) and current (A) are not correct, the device must not be connected to the
power supply.
2. Connect the IEC connector to the socket of the control box of the device.
3. Connect the grounding plug of the power supply cable to a correctly grounded and fused socket.
4. Make sure the power supply cable is not subjected to tensile or compressive force.
5. Mains connection: The mains socket must be easily identifiable by the user and freely accessible at
all times. The power cord plug is the all-pole disconnecting device from the mains supply.
Thermo Scientific
4
Start-up
Power supply connection
i160 / i250
S
TERI-CYCLE
4-15