Thermo Scientific STERI-CYCLE i160 User Manual page 138

Co2 incubator
Table of Contents

Advertisement

8
Cleaning and disinfection
Steri-run decontamination routine
6. Restart operation using auto-start as appropriate.
CAUTION Hot surfaces!
The handle and the screen of the glass door, the interior panel of the outer door as well as the
surfaces of the shelf system and of the workspace become extremely hot during the steri-run
decontamination routine.
During the routine run or immediately after completion of the run, always wear safety gloves when
touching these surfaces!
CAUTION Damage to the samples!
During the steri-run decontamination routine, the workspace is heated up to 180 °C. Make sure
that:
• all samples are removed from the workspace,
• all accessories are removed from the workspace.
Operating phases of steri-run decontamination:
The remaining run time of the steri-run decontamination routine describes the time between the start
or the current time status to the end of the cooling-downphase. The indicated remaining run times are
not measured values but merely used for orientation.
The routine is divided into three phases:
1. Heating phase,
2. decontamination phase,
3. cool-down.
Heating phase: approx. 2 h.
The workspace is heated up to 180 °C.
Decontamination phase: approx. 1.5 h.
After the decontamination temperature has been created, the decontamination phase of approx. 90
minutes is started. The temperature is maintained at 180 °C.
Cool-down phase: approx. 8 h.
The device cools down until the originally set temperature set value is reached.
i160 / i250
S
8-8
TERI-CYCLE
Thermo Scientific