Cleaning and disinfection
Wipe/spray disinfection
CAUTION Health hazard!
The surfaces of the workspace may be contaminated. Contact with contaminated cleaning liquids
may cause infections. Disinfectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety instructions and hygiene regulations!
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to protect your mucous membranes.
• Observe the safety instructions of the disinfectant's manufacturer and the hygiene supervisor.
Predisinfection:
1. Spray disinfectant onto the surfaces of the workspace and of the accessories and wipe the surfaces
clean.
CAUTION Moisture-sensitive components!
Do not spray the CO
2. Let the disinfectant work on the surfaces/internals as detailed in the manufacturer's instructions.
Removing accessories and shelf system:
1. Remove the shelves, then remove the entire shelf system from the workspace. The installation and
removal of the shelving system are described in the section
page
2. Remove the airbox and the HEPA-filter. The installation and removal of the airbox and the
HEPA-filter are described in section
i160 / i250
S
8-4
TERI-CYCLE
sensor and the
2
4-7.
-/N
sensor behind the air duct with disinfectant.
2
2
"Installing the shelf system"
"Replacing the HEPA-filter"
on
on
page
9-7.