Thermo Scientific STERI-CYCLE i160 User Manual page 66

Co2 incubator
Table of Contents

Advertisement

4
Start-up
Gas connection
The supplied gas pressure hoses feed the gas from the central gas supply system to the device.
Devices with combined CO
system are connected according to the connection diagram
O
/N
supply:
2
2
• Distributor connection for the O
Figure
pressure hose.
• Item
• Item
• Item
inlet filter (item
CO
supply:
2
• Item 4/: Distributor connection for the CO
devices. For looping the gas supply through, the devices must be interconnected with a gas pressure
hose.
• Item
• Item
• Item
inlet filter by a short gas pressure hose.
Note Looping the gas supply through:
By looping the gas supply through, up to 4 devices (independent of device type) can be supplied by
a CO
2
i160 / i250
S
4-14
TERI-CYCLE
/O
/N
2
2
4-13). For looping the gas supply through, the devices must be interconnected with a gas
2/Figure
4-13: Connect the gas container B to the connection of the gas monitoring system,
3/Figure
4-13: Connect the gas container A to the connection of the gas monitoring system.
1/Figure
4-13: The outlet of the gas monitoring system is connected at the factory to the gas
8/Figure
4-13) by a short gas pressure hose.
5/Figure
4-13: Connect the gas container B to the connection of the gas monitoring system,
6/Figure
4-13: Connect the gas container A to the connection of the gas monitoring system.
7/Figure
4-13: The outlet of the gas monitoring system is connected at the factory to the gas
gas supply or a O
-/N
gas supply.
2
2
connections that are equipped with the optional gas monitoring
2
(10/Figure
-/N
gas supply for up to three additional devices (hidden in
2
2
Figure 4-13
4-13).
gas supply for up to three additional
2
Thermo Scientific

Hide quick links:

Advertisement

Table of Contents
loading