Heating system
The workspace is heated by a surface heater system. The arrangement of the heating elements ensures
that condensate formation above the water reservoir is prevented.
The outer door of the unit and the perimeter of the door opening are heated as well. The heat radiated
onto the interior glass door/gas-tight screen prevents condensate formation.
The workspace of the device always remains visible, despite high humidity.
Rear panel openings
1
2
3
Figure 3-15. Rear panel openings
The access port with insert
(1/Figure
unit.
The pressure compensation opening
wall of the unit ensures pressure compensation between the workspace of the unit and the operating
room.
Thermo Scientific
(Figure 3-15
3-15) and allows for feeding cables, tubes or additional sensors into the workspace of the
Ø 42 mm / 1.66 inch) can be sealed with a plug
(3/Figure
3-15) located directly beneath the access port in the rear
3
Description of the device
Workspace Components
i160 / i250
S
TERI-CYCLE
3-21