Moisture-Sensitive Components; Cleaning Exterior Surfaces; Cleaning Operation Panels; Wipe/Spray Disinfection - Thermo Scientific 3962 Operating Manual

Environmental chamber
Table of Contents

Advertisement

| Cleaning and Disinfection
Chapter 8
|
8-4
Model 3962

Moisture-sensitive components!

Sensitive components!
Do not spray cleaning agent onto the controllers of the device. When wiping the device
clean, always make sure that moisture does not enter into these components.

Cleaning exterior surfaces

1. Thoroughly remove dirt residues and deposits using a solution of tepid water and dish
washing agent.
2. Wipe the surfaces clean using a clean cloth and clear water.
3. Then, wipe the surfaces dry using a clean cloth.

Cleaning operation panels

The operation panels are moisture-sensitive.
Do not spray or wipe with cleaner.
Clean using a dry cloth of 100% micro fibre.

Wipe/spray disinfection

The manual wipe/spray disinfection is carried out in three stages:
Predisinfection
Cleaning
Final disinfection
Alcoholic disinfectants!
Disinfectants having an alcohol content of more than 10% may form, in
combination with air, easily combustible and explosive gas mixtures.
When using such disinfectants, avoid open flames or exposure to excessive heat
during the entire disinfection process!
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned device
components thoroughly dry.
Observe safety regulations to avoid fire and explosion hazard caused by
alcohol-containing disinfectants.