| Cleaning and Disinfection
Chapter 9
|
9-4
Model 3907 & 3948

Cleaning Operation Panels

The operation panels are moisture-sensitive.
Do not spray or wipe with cleaner.
Clean using a dry cloth of 100% micro fibre.

Wipe / Spray Disinfection

The manual wipe/spray disinfection is carried out in three stages:

Predisinfection

Cleaning
Final disinfection
Alcoholic disinfectants
Disinfectants having an alcohol content of more than 10% may form, in
combination with air, easily combustible and explosive gas mixtures.
When using such disinfectants, avoid open flames or exposure to excessive heat during
the entire disinfection process!
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned device components
thoroughly dry.
Observe safety regulations to avoid fire and explosion hazard caused by
alcohol-containing disinfectants.
Chloride-containing disinfectants
Do not use chloride-containing disinfectants.
Predisinfection
Remove all samples from the work space and store them in a safe place.
Spray disinfectant onto the surfaces of the work space and of the accessories or wipe the
surfaces clean using disinfectant.