Chapter 4 Cleaning and disinfection

4.2 Wipe/spray disinfection

Caution
Alcoholic disinfectants!
Disinfectants having a alcohol content of more than 10% may form, in combination with air,
easily combustible and explosive gas mixtures.
When using such disinfectants, avoid open flames or exposure to excessive heat during the
entire disinfection process.
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned C
thoroughly dry.
Observe safety regulations to avoid fire and/or explosion hazard caused by alcohol-containing
disinfectants.
Caution
Chloride-containing disinfectants!
Do not use chloride-containing disinfectants.
Use only disinfectants that do not affect stainless steel.
Warning
Decontamination or cleaning agents!
Thermo Electron LED GmbH or his agent is to be consulted, if there is any doubt about the
compatibility of decontamination or cleaning agents.
1. Remove all samples from the work space and store them in a safe place.
2. Spray disinfectant onto the surfaces of the work space and of the accessories or wipe the
surfaces clean using disinfectant. Do not spray disinfectant onto the membrane filters.
3. Allow time for the disinfectant to act as specified by the manufacturer.
4. Wipe off the chamber surfaces and the work tray using lukewarm water mixed with standard
rinsing agents. Rinse completely using warm water.
5. Re-rinse the cleansed surfaces 3 to 5 times with autoclaved water to completely remove any
cleaning agent residues.
6. After this, wipe the cleansed surfaces, air baffle and internals dry with a soft, sterile cloth.
Cell Locker 50155007_C 27.05.21
10
ELL LOCKER