9
10 Move the retroreflector back to the starting (zero) point and check the beam
11 Secure the straightedge to the work surface with tape.
12 On the Linear screen, click on Reset. This establishes the current positions of

Make the Measurements

You are now ready to make your first measurements.
1
2
Keysight 5519A and 5519B User's Guide
If Strength bar is less than 60 percent, realign the optics by repeating steps 1
through 5. If this fails to increase the signal, follow these steps:
a
Remove the interferometer assembly. Using only the retroreflector, the
beam strength should be near 100 percent. If it is not, then:
-
Verify the laser turret is in the "OTHER" position — see Figure 13.
-
Check the retroreflector lens for oily films, fingerprints, or physical
damage.
-
If beam strength still is not close to 100 percent, realign the
retroreflector.
Check the interferometer assembly lenses for oily films, fingerprints, or
b
physical damage, and clean if necessary.
Slide the retroreflector along the straightedge approximately 25 mm (1 in)
away from the interferometer assembly. If the beam strength falls to
60 percent or below, realign the optics by repeating steps 1 through 5.
Move the retroreflector to the end of the straightedge (at least 75 mm; 3 in)
and check the beam strength again. Realign the optics if necessary.
strength again. If the beam strength is less than 60 percent, repeat steps 1
through 9.
the optics as the zero point for the measurement.
On the screen, click Start. This is your first measurement, at the zero (0)
position. The text in the top of the display will flash green while the
measurement is being taken.
Slide the retroreflector 25 mm (1 in) away from the interferometer
assembly along the straightedge. Be sure to maintain the signal strength.