To Prepare a Capillary Column for Installation

WARNI NG
C AUTIO N
Operation Manual

To Prepare a Capillary Column for Installation

Materials needed
• Capillary column
• Column cutter, ceramic (5181-8836) or diamond (5183-4620)
• Ferrules
0.27 mm id, for 0.10 mm id columns (5062-3518)
0.37 mm id, for 0.20 mm id columns (5062-3516)
0.40 mm id, for 0.25 mm id columns (5181-3323)
0.5 mm id, for 0.32 mm id columns (5062-3514)
0.8 mm id, for 0.53 mm id columns (5062-3512)
• Gloves, clean
Large (8650-0030)
Small (8650-0029)
• Inlet column nut (5181-8830 for Agilent 7890A)
• Loupe
• Septum (may be old, used inlet septum)
The GC operates at high temperatures. In order to avoid burns, do not
touch any parts of the GC until you are sure they are cool.
Procedure
Always wear clean gloves while handling any parts that go inside the
GC or analyzer chambers.
Installing GC Columns
2
29