To Remove the LTM Module Assembly from the Instrument

This procedure is used to remove an LTM column module for replacement or
for instrument storage.
Materials needed
• Gloves, clean
• Internal column nut (G2855-20530)
• Ferrules (SilTite)
• Wire for plug (G2855-60593)
• Union (G3182-60580)
• Safety glasses
• Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
Procedure
1 Load your maintenance method for cooling down the column enclosure. See
2 When the instrument state changes to Ready, turn off the inlet and the LTM
Be careful! The guard column heated enclosure or internal accessories may be hot
WA R N I N G
enough to cause burns. If either is hot, wear heat-resistant gloves to protect your
hands or allow the parts to cool before beginning the work.
3 Open the LTM door on the right side of the GC/MSD.
5975T LTM GC/MSD Operation Manual
Large (8650-0030)
Small (8650-0029)
0.3-mm id, for < 0.25 mm id column (5188-5361)
0.4-mm id, for < 0.32 mm id column (5188-5362)
"Column maintenance method"
column from the ChemStation.
Keep the analyzer under vacuum unless you are performing analyzer
maintenance.
Installing LTM Columns
on page 120.