4
Preparing the Pump

Capillary Coding Guide

Capillary Coding Guide
Syntax for Capillary Description
The tables below are your guide to identifying the proper specifications for your
capillary. On all capillaries, dimensions are noted in id (mm), length (mm) and,
where applicable, volume (µL). When you receive your capillary, these
abbreviations are printed on the packaging.
Using the guide: This fitting is coded as SPF, for Swagelok, PEEK, Fingertight.
Table 11
Type
The type gives some
indication on the primary
function, like a loop or a
connection capillary.
Key
Capillary
Loop
Seat
Tube
1260 Infinity II Flexible Pump User Manual
Capillary coding guide
Material
The material indicates
which raw material is
used.
Description
Key
Connection
ST
capillaries
Loop
Ti
capillaries
Autosampler
PK
needle seats
Tubing
FS/PK
Fitting left/fitting right
The fitting left/right
indicate which fitting is
used on both ends of the
capillary.
Description
Key
Stainless
W
steel
Titanium
S
PEEK
M
PEEK-coated
E
1
fused silica
Description
Swagelok +
0.8 mm Port id
Swagelok +
1.6 mm Port id
Metric M4 +
0.8 mm Port id
Metric M3 +
1.6 mm Port id