Using The Cuvette Holder - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector

Using the Cuvette Holder

When required:
Tools required:
Parts required:
This cuvette holder can be placed instead of a flow cell in the variable wavelength
detector. Standard cuvettes with standards in it, for example, National Institute of
Standards & Technology (NIST) holmium oxide solution standard, can be fixed in it.
This can be used for wavelength verifications.
94 1100 Series Variable Wavelength Detector Reference Manual
If your own standard should be used to checkout the
instrument.
None
Cuvette Holder G1314-60200
Cuvette with the "standard", e.g. NIST certified holmium oxide sample