Start The Calibration; Load The Plate Or Array Card; Start The Run; Review The Results And Complete The Calibration - Applied Biosystems ViiA 7 Real Quick Reference Manual

Time pcr system
Table of Contents

Advertisement

Calibrate the ViiA
7 System
2

Start the calibration

3
Load the plate or array
card
4

Start the run

5
Review the results and
complete the calibration
4
a.
Select the instrument to calibrate:
1. In the ViiA
7 Software, select Instrument Instrument Console. The console displays
instruments on the network.
2. Select an instrument icon, then click Add to My Instruments.
Note:
You must add the instrument to your instrument list before you can manage the
instrument.
b.
Start the calibration:
1. Select the instrument, then click Manage Instrument.
2. In the Instrument Manager, select the calibration, click Start Calibration, then perform the
calibration as instructed.
Load the calibration plate or array card into the instrument tray.
Confirm that the instrument has the correct plate adapter. Ensure that the plate or array card is
properly aligned in the holder:
• Load plates with the A1 position aligned with the A1 position of the tray.
• Load plates and array cards with the bar code facing the front of the instrument.
IMPORTANT!
Plates should be loaded and unloaded by trained operators who have been warned of
the moving parts hazard.
a.
In the Setup tab, select Please check the box when the calibration plate has been loaded,
then click Next.
b.
In the Run tab, click START RUN to start the calibration.
IMPORTANT!
Do not attempt to open the access door duing the run. The door is locked while the
ViiA
7 System is in operation.
a.
After the run, view the results in the Analysis tab to verify that the calibration run passed. If the
run failed, troubleshoot the failed calibration as described in the Applied Biosystems ViiA
Time PCR System User Guide.
b.
After you inspect results, click Next, then remove the plate or array card when the ViiA
System ejects the plate adapter.
WARNING!
PHYSICAL INJURY HAZARD. During instrument operation, the plate can be
heated to 100 °C. Before removing the plate, wait until it reaches room temperature.
c.
Click Finish to complete the calibration.
d.
Discard the array card; discard the plate, unless it can be reused. If you store the plate for reuse,
return it to the packaging sleeve, then return the packaged plate to the freezer.
Applied Biosystems ViiA
7 Real-Time PCR System Quick Reference Guide