Performing the Amplification Run

Performing the Amplification Run
Select the Instrument tab.
1.
By default, the standard PCR conditions for the
PCR step are displayed.
®
AmpErase
UNG Activation
HOLD
2 min @ 50 ° C
Accept default values for:
2.
• Sample volume – 50 µL (20 µL for the Fast
system).
• 9600 Emulation – selected.
The 9600 Emulation feature is not
Note:
available for the 7300 instrument.
Notes
32
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Plus/Minus Assay Getting Started Guide
Times and Temperatures
Initial Steps
®
AmpliTaq Gold
DNA
Polymerase Activation
HOLD
10 min @ 95 ° C
PCR (Each of 40 cycles)
Melt
CYCLE
15 sec @ 95 ° C
Anneal/Extend
CYCLE
1 min @ 60 ° C