Tsplink.master - Keithley 2600B Reference Manual

System sourcemeter instrument
Hide thumbs Also See for 2600B:
Table of Contents

Advertisement

Series 2600B System SourceMeter® Instrument Reference Manual

tsplink.master

This attribute reads the node number assigned to the master node. This attribute is not available on the Models
2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
masterNodeNumber = tsplink.master
masterNodeNumber
Details
After doing a TSP-Link reset (tsplink.reset()), use this attribute to access the node number of
the master in a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Also see
tsplink.reset()
tsplink.node
This attribute defines the node number. This attribute is not available on the Models 2604B/2614B/2634B.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
nodeNumber
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until tsplink.reset() from an earlier TSP-Link
instrument is executed on any node in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
tsplink.node = 3
2600BS-901-01 Rev. C / August 2016
Affected by
Not applicable
The node number of the master node
(on page 7-385)
Affected by
Not applicable
The node number of the instrument or enclosure (1 to 64)
Section 7: TSP command reference
Where saved
Default value
Not applicable
Not applicable
Where saved
Default value
Nonvolatile
1
memory
Sets the TSP-Link node for this instrument to
number 3.
Store the TSP-Link master
node number in a variable
called LinkMaster.
7-383