Model 2657A High Power System SourceMeter® Instrument Reference Manual

tsplink.master

This attribute reads the node number assigned to the master node.
Type
TSP-Link accessible
Attribute (R)
Yes
Usage
masterNodeNumber = tsplink.master
masterNodeNumber
Details
After doing a TSP-Link reset (tsplink.reset()), use this attribute to access the node number of the master in
a set of instruments connected over TSP-Link.
Example
LinkMaster = tsplink.master
Also see
tsplink.reset()

tsplink.node

This attribute defines the node number.
Type
TSP-Link accessible
Attribute (RW)
Yes
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
nodeNumber
Details
This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until the next time tsplink.reset() is executed on any node
in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
tsplink.node = 2
Also see
tsplink.reset()
tsplink.state
2657A-901-01 Rev. B/December 2012
Affected by
Not applicable
The node number of the master node
(on page 7-347)
Affected by
Not applicable
Set node to a number (1 to 64)
(on page 7-347)
(on page 7-347)
Section 7: TSP command reference
Where saved
Default value
Not applicable
Not applicable
Where saved
Default value
Nonvolatile
1
memory
Sets the TSP-Link node to number 2.
Store the TSP-Link master
node number in a variable
called LinkMaster.
7-345