Troubleshooting/Errors

3800 EEPROM storage error

A problem has been identified with the User Method storage device.
The method parameters have failed to store, or were retrieved with
error and hence ignored.

5004 Signals not increasing

The calibration has failed. The signal readings of the calibration
standard solutions are not increasing in the expected order. The
instrument cannot calibrate using this sequence of standards.
Suggested actions:
Move the cursor to the last standard measured and press Enter to
clear the field and remove the invalid concentration from the
calibration. If more than one standard has been measured the
calibration should now be valid, indicated by the symbol Cal in the
bottom right of the display screen area.
Alternatively, remeasure the previous standard (ensuring you are
using the correct standard) by returning the cursor to the
appropriate field and pressing Read.

5005 Slope test failure

The calibration has failed. The slope of the calibration curve changes
too much between successive standard signal readings.

5006 Calibration fit failure

The calibration has failed. A calibration curve could not be fitted to
the measured standards. Effective calibration is indicated by the
symbol Cal in the bottom right of the display screen area.

5008 Reslope signal out of range

Reslope calibration has failed. The reslope standard signal reading
was too far out to be used. (The Reslope signal value cannot differ
from the original value by more than 25%.)