To set up MassHunter program for the Nanospray Ion Source

1 In the MassHunter Acquisition program, make sure that the source type is
2 Set the drying gas flow and temperature parameters. See
Table 1
I-Funnel instruments with Hexabore capillaries
Non I-Funnel instruments with single bore capillaries
3 Set the VCap voltage to 900 volts.
Nanospray Ion Source User Guide
set to HPLC-Chip in the Tune Panel.
Operating the source with parameters above these set points can result in
reduced performance, and can lead to frequent needle clogs.
Maximum drying gas temperature and flow rate
Installation
Table
Maximum Drying
Maximum Flow
Gas Temperature
Rate
225°C
15 L/minute
365°C
8 L/minute
1