Temperature Calibration - Thermo Scientific AquaSensors RDO Pro-X Manual

Optical dissolved oxygen measurement system
Table of Contents

Advertisement

Section 7
Calibration and Configuration
|
34
RDO Pro-X Optical Dissolved Oxygen Measurement System
|
Once the actual 1-point sample calibration has been initiated; the display will indicate that a
calculation is progressing. During this calculation it is important that nothing disturb the
measurement sample.
To abort the calibration during this calculation, press ESC.
1-Point Sample
Calibrating...
When the calibration calculation is complete, the result will be displayed on the second line of
the display. If the calibration is successful, the message will be CAL OK. If the calibration is not
successful, the message will indicate the reason for the failure. In either case, press ENTER to
continue.
1-Point Sample
Cal OK
MENU
ESC
If the calibration is successful, pressing ENTER will return to the measure screen. If the
calibration is not successful, pressing ENTER will return to the beginning of the 1-point sample
calibration. Reasons for calibration failure include a sample that is not in the correct
measurement range or is changing in value too quickly.

Temperature Calibration

Press the ENTER key from the TEMPERATURE calibration menu to initiate the calibration
process.
≡Calibrate
►Temperature
MENU
ESC
ENTER
ENTER
Thermo Scientific AquaSensors

Advertisement

Table of Contents
loading

Table of Contents