4.7

Temperature Calibration

All DataStick sensor heads (transducers) measure temperature the same way and share a common calibration process.
Select TEMPERATURE from the calibration menu to begin a temperature calibration. The status box will prompt the user to enter
the desired calibration value.
Once the desired value for calibration is entered, the CALIBRATE button becomes active. Once the displayed temperature value is stable
and near expected values, click the CALIBRATE button to initiate the calibration or click EXIT to return to the main measure screen.
Once the calibration has been initiated it will take about 30 seconds depending on temperature measurement stability. During this period,
the abort sequence can be initiated as shown with the active Abort button.
Thermo Scientific AquaSensors™ DataStick™ Measurement System User Guide
93