Dut Connections - Keithley 6430 Quick Results Manual

Sub-femtoamp remote sourcemeter
Hide thumbs Also See for 6430:
Table of Contents

Advertisement

Figure 5
Basic input-output configuration
Center Conductor
A) Remote PreAmp Triax Connector Configuration
HI (center conductor)

DUT connections

2-wire sense connections are shown in Figure 6, and 4-wire sense connections are shown
in Figure 7. Use 4-wire remote sensing for the following source-measure conditions:
The error caused by test lead resistance with 2-wire sensing is not acceptable.
Optimum ohms, V-source, and/or V-measure accuracy is required.
Inner Shield
Outer Shield
6430 Mainframe
HI
INPUT/OUTPUT
LO
Zener
Clamp
Chassis
Ground
Triax to Alligator Clip
Guard (inner shield)
LO (outer shell)
B) Terminal identification for 6430-322-1A cable
C) Mainframe Input/Output Configuration
Quick Results Guide
HI (In/Out or Sense)
GUARD (Cable)
LO (In/Out)
42V Peak DC Max
or
10.5mA Max
Chassis Ground Screw
(on Rear Panel)
Red boot
Black boot
Green boot
7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents