Multiposition stream selection valve with sampling valve

Advanced Operation Manual
Load time Time in minutes that the valve remains in the
Load position before becoming ready.
Inject time Time in minutes that the valve remains in the
Inject position before returning to the Load position.
The sampling valve cycle is:
The sampling valve rotates to the Load position. Load time
1
begins. Valve is not ready.
Load time ends. The valve becomes ready.
2
If everything else is ready, the GC becomes ready. If
3
anything is not ready:
If you are using Clock Table or sequence control, the
GC waits until everything is ready, then executes the
valve inject command.
If you are not using Clock Table or sequence control,
the valve injection can be made at any time from the
keyboard.
The sampling valve rotates (keyboard command or
4
sequence control) to the Inject position. Inject time begins.
The run begins.
Inject time ends. Return to step 1.
5
Several manufacturers provide multiposition stream selection
valves that can be driven by valve drivers 1 through 4. Only
one multiposition valve can be configured.
If a valve is configured as a multiposition valve and has a
BCD position output connected to the GC, the valve position
can be selected directly.
6
Valves
223