5-6 Recovering calibration ROM data
If an abnormality occurs in the sensor's calibration ROM, you can recover the backed up calibration ROM data
into the sensor.
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.61
Explorer pane
Edit pane
2
Click the [ROM recovery] in [Sensor setup].
Select the calibration ROM backup data file.
Important
The calibration ROM data is different for each sensor serial number. Select the backup file that matches the sensor
serial number. Measurement will not be correct unless they match.
ZW User's Manual
: [Device Group] | [(Sensor Name)] (double click)
: [Tool] icon (
)
Recovering calibration ROM data
133