Select the operation during non-
measurement at [Non-Measurement
Setting].
Note
The operations for when measurement is not possible can also be set with key operations on the Sensor Controller.
Setting operation when measurement is not possible p.344
Setting the Clamp Value
Set the clamp value to output when "Clamp" is selected as processing when measurement is not possible.
Item
Setting item
Analog output
Clamp output
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.61
 Explorer pane
 Edit pane
 I/O Setting Screen
2
Select the output value at [Output value
when clamp].
ZW User's Manual
Setting value
When the analog output destination is set to current
MIN (approx. 3.4mA)/4 to 20mA (1mA increments)/
MAX (default value: approx. 21mA)
When the analog output destination is set to voltage
MIN (approx. -10.8V)/-10 to 10V (1V increments)/
MAX (default value: approx. 10.8V)
: [Bank] | [(Bank Data Name)] (double click)
: [I/O Settings] icon (
: [Analog output]
Description
Select the current/voltage at clamp
output.
)
Parallel I/O connection