Recovering Calibration Rom Data - Omron ZW-XC7 User Manual

Confocal fiber type displacement sensor
Table of Contents

Advertisement

5-8 Recovering calibration ROM data

If an abnormality occurs in the sensor's calibration ROM, you can recover the backed up calibration ROM data
into the sensor.
1
Set the operating mode to the FUNC mode.
3-3 Switching operation modes p.85
 Explorer pane
→ Edit pane
2
Click the [ROM recovery] in [Sensor setup].
Select the calibration ROM backup data file.
Important
The calibration ROM data is different for each sensor serial number. Select the backup file that matches the sensor
serial number. Measurement will not be correct unless they match.
ZW-7000 User's Manual
: [Device Group] | [(Sensor Name)] (double click)
: [Tool] icon (
)

Recovering calibration ROM data

159

Advertisement

Table of Contents

Troubleshooting

loading

This manual is also suitable for:

Zw-7000 series

Table of Contents