Omron ZW-C1*AT User Manual page 149

Zw series displacement sensor confocal fiber type displacement sensor
Table of Contents

Advertisement

Item
Setting item
Judgment
Timer mode
output
Timer time
1
Set the operating mode to the FUN mode.
3-3 Switching operation modes p.61
 Explorer pane
 Edit pane
 I/O Setting Screen
2
Set [Hysteresis Width].
3
Select the judgment output timing to match
operation of the external device from [Timer
settings] - [Timer mode].
4
Sets [Timer time].
Note
The operations for judgment output can also be set with key operations on the Sensor Controller.
Setting Operation at Judgment Output p.343
ZW User's Manual
Setting value
Description
One Shot
When the judgment result changes to PASS, the PASS output is executed
for the time set to [Timer Duration].
1 (default value) to
Sets the timer duration when the timer mode is other than OFF.
5000 [ms]
: [Bank] | [(Bank Data Name)] (double click)
: [I/O Settings] icon (
: [Judgment]
Measured value
HIGH threshold
value
LOW threshold
value
ON
HIGH output
OFF
ON
PASS output
OFF
ON
LOW output
OFF
)
: Timer time
Parallel I/O connection
147