Emptying The Diva Metering Module - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Emptying the DIVA metering module

1 Hold empty anesthetic agent bottle (A) under
the drain opening on the bottom of the DIVA me-
tering module.
2 Swivel the lever slowly.
3 Pull out the locking bolt and fold it down.
4 Use a screwdriver for hexagon sockets to open
the drain plug (B).
5 Close the drain plug again after emptying.
6 Insert the locking bolt again and press in up to
the stop.
7 Close the lever.
8 Close the anesthetic agent bottle.
Instructions for use Zeus Infinity Empowered SW 2.n
B
A
Assembly and preparation
83

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents