End of operation

Switching Zeus IE into system standby
System standby?
B
1 Touch the Start/Standby button (A).
All settings previously modified in operation are
preserved.
The System standby? dialog window will be
opened.
2 Confirm Stop ventilation? button (B) with ro-
tary knob.
Or continue ventilation:
Close dialog window.
Pressing the rotary knob automatically switches the
device to the system standby screen.
Standby
156
System standby is the standby mode in which:
– Fresh gas is switched off
– Quick start dialog window appears
– Electrical power consumption is minimal
– Operation can be resumed quickly
– The device can be switched off