Autoset Limits - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Alarms

Autoset limits

Alarms
B
The Autoset button (A) is used to adjust the alarm
limits displayed in the Alarm limits dialog window
(B) to current measured or set values.
WARNING
The Autoset function always refers only to the
alarm limits highlighted in yellow on the par-
ticular page!
This resets the previously active alarm limits!
Disabled alarms are not automatically ena-
bled.
Autoset can be performed for the following areas:
– Noninvasive
– Invasive
– Ventilation
– Anesth.
1 Touch the corresponding tab, e.g., (C).
2 Touch the Autoset button (A). The Autoset but-
ton is preselected in yellow. Alarm limits that will
be recalculated by the Autoset function are
highlighted in yellow on the respective page.
3 Touch the Autoset (A) button. The alarm limits
highlighted in yellow will be recalculated.
The limit is adjusted based on the formula:
Upper limit
=
Current measurement + X %
Lower limit
=
Current measurement – X %
172
For exceptions, see table "Adjusting the lower and
upper limit values according to set points and pa-
tient demographic data" on page 174.
C
The Autoset difference (X) is set in the Autoset
limits dialog window.
Touch the Autoset limits tab (D).
Alarms
A
The Autoset difference (X) can be set for the follow-
ing areas:
E Hemodyn.
F IBP
G Anesth./ Ventilation
Instructions for use Zeus Infinity Empowered SW 2.n
D
E
F
G