Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Warning
180
Apnea
Caution
250
Apnea
Note
050
Apnea alarms off
Warning
180
Apnea CO
Caution
250
Apnea CO
Note
175
Apnea flow
Warning
180
Apnea flow
Caution
250
Apnea flow
258
Cause
No automatic ventilation. No
breath detected for 30 sec.
No spont/ass. ventilation: No
breath detected for 60 sec.
No automatic ventilation: No
breath detected for 15 sec.
No spont/ass. ventilation: No
breath detected for 30 sec.
Apnea alarms are sus-
pended. To be used only dur-
ing intubation, suctioning
and manual ventilation.
No automatic ventilation: No
2
CO
breath detected for
2
30 sec. No spont/ass. venti-
lation: No breath detected by
CO
for 60 sec.
2
No automatic ventilation: No
2
CO
breath detected for
2
15 sec. No spont/ass. venti-
lation: No breath detected by
CO
for 30 sec.
2
No sufficient flow dynamics.
Flow sensor not calibrated.
No automatic ventilation: No
flow breath detected for
30 sec. No spont/ass. venti-
lation: No breath detected by
flow for 60 sec.
No automatic ventilation: No
flow breath detected for
15 sec. No spont/ass. venti-
lation: No breath detected by
flow for 30 sec.
Remedy
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Please pay special atten-
tion! To be used only during
intubation and suctioning
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
To check and calibrate, per-
form self test or leak test.
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!
Ventilate patient by hand
immediately! Check tubes
for disconnection! Check
ability to breath spontane-
ously!