Disabling Options; More Settings - Dräger Zeus Infinity Empowered Instructions For Use Manual

Hide thumbs Also See for Zeus Infinity Empowered:
Table of Contents

Advertisement

Configuration
The Set default button (I) resets MediBus to the
default values:
– Port: Off
– Baud rate: 9600
– Parity: Even
– Stop bits: 1
– MIB/IDS filter: On
– Commands to MediBus device: On (COM 1)
– MediBus version: SW 4.n
Check the following device settings in the De-
vice identification list:
– Device number
– Device name
– Device version
– MediBus version
Check the following device settings in the
Linked device list:
– Device number
– Device name
– Device version
– MediBus version
Status: not connected or connected

Disabling options

The page Basic settings > Disable options
opens.
Basic settings
A
B
C
D
E
F
L
Enabled options can be disabled and again ena-
bled. The button On is preset. If an option is gener-
ally not enabled, the On button is grayed out and
the Off button enabled.
348
Changes take effect only after activating the Adopt
Restart button (L). The Cancel button (M) cancels
the process.
The following options can be switched off:
A Patient monitor
B ST segment analysis
C Display waveforms on screen 2
D IVenus - remote control
E Ext. fresh-gas outlet
F Extended arrhythmia
G Infinity network
H Web access 1
I
J Web access 3
K Web access 4

More settings

The following settings can be configured in the Ba-
sic settings > More settings dialog window:
– Gas prices
– Ventilation settings (ventilation settings)
– Ventilation modes (ventilation modes)
– VA/N
– Calculat. settings (calculation settings)
– Total consumpt.
G
H
I
J
K
M
Instructions for use Zeus Infinity Empowered SW 2.n
Web access 2
O (volatile anesthetic agent/N
2
O)
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents