Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Caution
100
ART M low
Note
150
ART nonstatic
Caution
050
ART out of range
Caution
125
ART S high
Caution
130
ART S low
Warning
250
ART static
Note
050
ART transducer fail
260
Cause
Invasive pressure value out-
side alarm limits.
Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Invasive pressure signal out
of measurement range.
Invasive pressure value out-
side alarm limits.
Invasive pressure value out-
side alarm limits.
Static pressure: -No pulse,
-Press. transducer line
blocked, -Catheter tip lodged
against vessel wall, -Clot on
the catheter tip.
Pressure transducer failed.
Invasive pressure cannot be
measured.
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.
Check patient and take ap-
propriate action. Access the
Pressure Labels menu and
assign correct label. Check
equipment and replace if
necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Open sys-
tem by turning the 3-way
tap. Follow procedure for
dislodging catheters. Fol-
low procedure for clotted
catheters.
Check transducer and con-
nections. Exchange trans-
ducer.