Cleaning, disinfection, and sterilization

Uncritical medical devices

Manual disinfection with simultaneous cleaning
When selecting a suitable disinfectant, adhere to
the country-specific lists of disinfectants. The list of
the German Association for Applied Hygiene (Ver-
bund für Angewandte Hygiene VAH) applies in Ger-
man-speaking countries.
Strictly observe the manufacturer's specifications
on the disinfectants. Manufacturers may change
the composition of disinfectants over time.
Procedures:
1 Remove soiling immediately with a cloth soaked
in disinfectant.
WARNING
Risk of electric shock or device failure
Liquid that enters into the device can cause
the device to malfunction or may damage the
device and endanger the patient.
Only disinfect the device surface and cables
using the moist scrub-wipe disinfection
method and ensure that no fluid enters the de-
vice.
2 Perform surface disinfection by scrubbing and
wiping.
3 Remove disinfectant residues after the contact
time has elapsed.
368

Semicritical medical devices

Manual cleaning
Perform manual cleaning preferably under running
water and with commercially available cleaning
agents (pH ≤12).
Procedures:
1 Wash off surface soiling under flowing water.
2 Use cleaning agents in accordance with the
manufacturer's instructions. Make sure that all
surfaces and interior spaces to be cleaned can
be reached. Use suitable brushes if necessary.
3 Thoroughly rinse components under running
water until cleaning agent residues are no
longer discernible.
4 Inspect components for visible soiling and dam-
age. Repeat manual cleaning if necessary.