Fault – Cause – Remedy
Classifi-
Pri-
Alarm text
cation
ority
Note
100
KVO pump 4
Note
050
LA check zero
Note
020
LA electrode fail
Note
025
LA hardware fail
Caution
090
LA M high
Caution
100
LA M low
Note
150
LA nonstatic
286
Cause
Volume limit of infusion 4 has
been reached. 'Keep vein
open' function is activated.
Infusion continues with a low
infusion rate (1 mL/h).
Pressure transducer for
specified parameter requires
zeroing. The blood pressure
zero of monitor may not cor-
respond to the peripheral de-
vice.
Lead-off condition for indi-
cated lead detected: -Broken
cable - Loose lead wire
- Faulty lead wire - Wrong
lead -Dried out gel -Bad con-
tact.
Invasive blood pressure
hardware failure.
Invasive pressure value out-
side alarm limits.
Invasive pressure value out-
side alarm limits.
Press. transducer zeroing
failed because of: Excessive
signal noise A non-static
waveform
Instructions for use Zeus Infinity Empowered SW 2.n
Remedy
Check infusion. Check vol-
ume alarm limit. To change
KVO configuration at the
pump refer to the module
DPS instructions.
Zero the transducer.
Check defective cables and
electrodes. Reapply gel to
electrodes. Make sure the
skin is properly prepped.
Select another ECG/ST
lead. Set correct type of ca-
ble.
Check hardware and re-
place. Call DrägerService.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Check patient and take ap-
propriate action. Change
the alarm limits if required.
Check equipment and re-
place if necessary.
Keep all tubing motionless,
then rezero. Change the
transducer. Check 3-way
tap, then rezero.